4-8 | Dycor CG1000-RTP Oxygen Analyzer
Analog Range Key
All functions in this section are directed to a specific analog output port.
If a system password is enabled, you will must correctly enter this
password before accessing
Analog Range
key functions.
Analog Key Menu
Function/Options
set current Range
Define
analog output range as 0-to-20 mA or 4-to-20
mA (
Set Current Mode
).
4% o
2
= 4 mA
10% o
2
= 20 mA
set Track/Hold
Define information sent to analog output port during a
calibration.
decide whether to hold the output at the last process
reading taken before that start of calibration oR track
the calibration reading during cal or verify.
set current Mode
Define the analog output range as:
4 to 20 mA; 0 to 20 mA
output Filtering
Decide whether to enable output filtering. If enabled,
decide how much to smooth out abrupt changes in
readings.
select Function
Define the type of information the analog output should
track.
o
2
cell Temperature
Thermocouple Millivolts
cell Millivolts
NOTE
NOTE