A-8 | Dycor CG1000-RTP Oxygen Analyzer
Figure A-3.
Main variable table (2).
03
iout_flags
HEX
Analog output Flags
XX XX
1 2
Byte 1 (bit#)
- Analog output #2
7 - Future use
6 - 0 = (4-20) ma
1 = (0-20) ma
5 - 0 = Holds during verify
1 = Tracks during verify
4 - 0 = Holds during cal
1 = Tracks during cal
- Analog output #1
3 - Future use
2 - 0 = (4-20) ma
1 = (0-20) ma
1 - 0 = Holds during verify
1 = Tracks during verify
0 - 0 = Holds during cal
1 = Tracks during cal
Byte 2 (bit#)
- Analog output #2
7 - Future use
6 - 0 = (4-20) ma
1 = (0-20) ma
5 - 0 = Holds during verify
1 = Tracks during verify
4 - 0 = Holds during cal
1 = Tracks during cal
- Analog output #1
3 - Future use
2 - 0 = (4-20) ma
1 = (0-20) ma
1 - 0 = Holds during verify
1 = Tracks during verify
0 - 0 = Holds during cal
1 = Tracks during cal
04
Probe type
inT
same as command letter 05
05
display line #1
inT
0 - Blank line
2 - oxygen
4 - cell Temperature
6 - cell Millivolts
10 - current Time and date
12 - user Text
14 - Flow
06
display line #2
inT
same as command letter 05
07
display line #3
inT
same as command letter 05
08
oxygen
FloAT
0B
cell Temp
FloAT
0c
cell mv
FloAT
0d
T/c mv
FloAT
COMMAND
VARIABLE NAME
FORMAT
DATA FORMAT
LETTER
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...