4-6 | Dycor CG1000-RTP Oxygen Analyzer
Upon completion of the primary calibration, the system notifies you
that a standard calibration is required by displaying a
Calibration Re-
quired
message. Oxygen readings on the display will not be accurate
until you perform a regular calibration.
A 15-minute recovery time follows the primary calibration to allow
the cell to stabilize. Wait an additional 45 minutes, then perform a
regular calibration.
Communication
Define RS-232 communications parameters between the CG1000-RTP and a
host device.
Baud Rate
Define the baud rate that the analyzer and the host comput
-
er use to communicate. choices are 300, 600, 1200, 2400,
4800, and 9600 baud.
Node Address
Define the physical address for the control unit (between 0
and 255) on a network so the host computer can identify the
control unit if using network communications.
Enable Serial
Temporarily disable cg1000-RTP serial communications.
Carrier Gas
Define which carrier gas is used to deliver the calibration gases to the analyzer
and calibrate the analyzer flow sensor to improve its accuracy.
You must select the proper carrier gas to ensure proper calibration of
the analyzer.
Define carrier gas
1. From the
Setup
key menu, select
Carrier Gas
.
2. Select the
Carrier Gas
sub-menu.
3. Select the carrier gas used to deliver gases to the analyzer. Choices are
nitrogen, carbon dioxide, argon, or helium. If measuring close to 100%
oxygen, we recommend that you select nitrogen as the carrier gas.
NOTE
See the “Serial Communica-
tions” appendix for complete
information on serial com-
munications.
NOTE
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...