3-2 | Dycor CG1000-RTP Oxygen Analyzer
Mechanical Installation
Sample Gas
Requirements
sample gas/calibration gas set-up
PPM Measurement suggestions
Sensor
Ambient Temperature Range
-0 °c to 40 °c (32 °F to 104 °F)
Sample Flow Rate
150 cc/min recommended
Max Sample Flow Rate
200 cc/min
Max Sample Pressure Limit
3 Psig
Max Sample Inlet Temperature
700°c (160°F)
Locate the CG1000-RTP as close to sampling system as possible.
The accuracy of the oxygen readings are pressure- and flow-dependent.
Maintain the same
flow rate and
inlet pressure during both calibration
and normal operations. Keep the flow rate between 125 to 175 cc/min
to maintain accurate readings. To detect when the flow exceeds these
limits, Alarm 3 can be programmed to alarm on both the high and low
flow rate.
Figure 3-1 shows the sample inlet and outlet ports on the CG1000-RTP.
The sample gas should be clean and dry. If the sample contains any con-
densing moisture, the entire sample line should be heat-traced.
Be aware that combustibles in the sample can burn with available
oxygen to provide lower-than-expected oxygen levels, especially if
measuring oxygen in the low PPM ranges.
Keep the inlet pressure as low as possible to prevent damage to the cell or
cell O-ring seal. The sample gas inlet pressure into the CG1000-RTP must
be less than 3 PSIG.
Figure 3-1. Rear of CG1000-RTP analyzer.
NOTE
If the sample contains
particulates greater than 15
microns, install a filter on the
sample gas inlet.
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...