Configuration and Calibration
43
3.7.4
Configuration Parameters (Continued)
Option Menu
Comm Port
(F7)
Allows selection of any available Communications Port,
Polling Address and Polling Address for HART multidrop
Dropdown boxes for:
Comm Port: All
available ports
Polling: 0 – 15
(Default = 0)
Demo Mode
Allows selection of Drexelbrook instrument type for
simulation.
Check type of
Drexelbrook
Instrument to
simulate
Run Demo
Initiates simulation of selected Drexelbrook instrument
Figure 3.7-5
3.7.5
Configuration examples (HART/4 – 20 mA output):
All units are factory configured to read level referenced to the bottom of
the sensor and it is not typically required to edit the configuration.
If additional configuration is required use the following examples.
Fig. 3.7-5
is for level and volume or
Fig. 3.7-6
is for distance. Configuration
can be performed either with HRTWin Software or via the display.
Level Configuration:
• Level Units: Inches
• Tank Height: 120
• Sensor Offset: 6
• LRV: 0
• URV: 110
• Material Type: Oil
Содержание Impulse GW Series
Страница 4: ...Contents...
Страница 6: ...Section 1...
Страница 16: ...Section 2...
Страница 25: ...Installation 19 2 6 Dual Compartment Housing...
Страница 26: ...Section 3...
Страница 56: ...Section 4...
Страница 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Страница 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Страница 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Страница 86: ...Section 6...
Страница 92: ...Section 7...
Страница 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Страница 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Страница 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Страница 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...