Configuration and Calibration
45
Figure 3.7-7
Sensor Offset
3.7.6 Level Configuration
Sensor Offset must be applied in cases where:
• A Volume Output is required and the Sensor Mounting Reference
extends below the top of the tank or, the Sensor Mounting Reference
is mounted above the top of the tank.
See Figure 3.7-7
• If the Sensor Mounting Reference is mounted above the top of the
vessel, a positive value is entered in Sensor Offset .
• If the Sensor Mounting Reference is mounted below the top of the
tank, a negative value is entered in
Sensor Offset
.
4.
Click on
Write to Transmitter
.
122"
Sensor offset
-13"
Transmitter
Transmitter
4" dia. nozzle
4"
122"
Sensor offset
Positive Sensor Offset
Negative Sensor Offset