Configuration and Calibration
63
4.7 Data Points
The following data points are available in read-only:
Data point
Notes
Product level
Product level in engineering units selected
System status
0 = SEARCHING FOR SIGNAL, 1 = TRACKING on target, 2 = NEAR
ZONE, 3 = LOST SIGNAL/EMPTY PROBE
Probe length
Length from tip of sensor to mounting reference point
Tank height
Tank height in engineering units selected
Current highest signal
The highest peak signal of product level
Current max signal area The maximum signal area of product level
Product distance
Distance from the bottom of the sensor process connection to
product level
Product volume
In volume units selected
Level/distance units
44 = Ft, 45 = Meters, 47 = inches, 48 = CM, 49 = MM
Volume units
40 = Gallons, 41 = Liters, 42 = Imperial gallons, 43 = cubic meters,
46 = Barrels
TDR material type
0 = CUSTOM, 1 = WATER, 2 = OIL
Sensor offset
Used if sensor is located above or below the tank height value
Upper dead zone
The top portion of the sensor where level can't be measured
Lower dead zone
The bottom portion of the sensor where level can't be measured
Damping
Time in seconds to dampen out variation in signal due to agitation or
rapid level change
Max volume capacity
max volume in engineering units selected
Vessel type
0 = custom vessel, 1 = vertical, 2 = HCYL flat ends, 3 = HCYL
dished ends, 4 = HCYL hemisphere, 5 = spherical
Cable type
0 = SINGLE CABLE, 1 = TWIN CABLE, 2 = SINGLE ROD, 3 = TWIN
ROD, 4 = COAX
Threshold value
Preset signal threshold of hydrocarbon (1200) or aqueous loquids
(1500)
Signal area threshold
Threshold signal area
Signal being sampled
0 = AMP OUT, 1 = PRE AMP OUT
Date code month
1 thru 12
Datecode day
1 thru 31
Date code year
4 digit year
Содержание Impulse GW Series
Страница 4: ...Contents...
Страница 6: ...Section 1...
Страница 16: ...Section 2...
Страница 25: ...Installation 19 2 6 Dual Compartment Housing...
Страница 26: ...Section 3...
Страница 56: ...Section 4...
Страница 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Страница 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Страница 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Страница 86: ...Section 6...
Страница 92: ...Section 7...
Страница 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Страница 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Страница 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Страница 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...