5.1.4 Standards and approvals
In Compliance with 2014/34/EU ATEX Directive. It is certified for use in hazardous
areas by FM certification agency under FM10ATEX0029X
Intrinsically Safe Apparatus Certification – FM10ATEX0029X
Impulse T130a/GWe Series 2-wire TDR Level Measurement System
II 1G Ex ia IIB T4, Ga; Ta = -30˚C to 70˚C ;; IP66/ IP67
II 2D Ex ia tb IIIC T90°C Db; Ta = -30˚C to 70˚C; IP66/ IP67
Entity Parameters or Energy Limitation Parameters :
Ui = 30 V, Ii = 140 mA, Pi = 1 W, Ci = 0, Li = 0
a = 0 or 1 (Electrical Connection)
e = 16 character numbering system which does not affect safety
Flame Proof /Intrinsically Safe Apparatus Certification FM10ATEX0029X
Impulse T140a/GWe Series 2-wire TDR Level Measurement System
II 2G Ex d ia IIB; T4, Gb; Ta = -30˚C to 70˚C; IP66/ IP67
II 2D Ex ia tb IIIC; T90°C Db Ta = -30˚C to 70˚C; IP66/ IP67
a = 0 or 1 (Electrical Connection)
e = 16 character numbering system which does not affect safety
Specific Conditions of Use
Impulse Ta30b/GWe Series 2-wire TDR Level Measurement System
1. The apparatus enclosure contains aluminum and is considered to constitute a
potential risk of ignition by impact or friction. Care must be taken during installation
and use to prevent impact or friction.
Impulse Ta40b/GWe Series 2-wire TDR Level Measurement System
1. The apparatus enclosure contains aluminum and is considered to constitute
a
potential risk of ignition by impact or friction. Care must be taken during
installation and use to prevent impact or friction.
2. Consult the manufacturer if dimensional information on the flameproof
joints is necessary.
5.1.5 Warning
Read the ATEX approval certificate for the Impulse carefully and observe
boundary conditions given.
Содержание Impulse GW Series
Страница 4: ...Contents...
Страница 6: ...Section 1...
Страница 16: ...Section 2...
Страница 25: ...Installation 19 2 6 Dual Compartment Housing...
Страница 26: ...Section 3...
Страница 56: ...Section 4...
Страница 82: ...Impulse Series 76 5 2 1 System Dimensions Continued...
Страница 83: ...Supplementary ATEX Installation Instructions 77 5 2 1 System Dimensions Continued...
Страница 84: ...Impulse Series 78 5 2 1 System Dimensions Continued...
Страница 86: ...Section 6...
Страница 92: ...Section 7...
Страница 93: ...Control Drawings 87 Section 7 Control Drawings 7 1 FM FMC Control Drawings...
Страница 94: ...Impulse Series 88 7 1 FM FMC Control Drawings Continued...
Страница 95: ...Control Drawings 89 7 2 ATEX Control Drawings...
Страница 96: ...Impulse Series 90 7 2 ATEX Control Drawings Continued...