The table below provides a summary of the Model 430 Programmer
setup parameters for this example.
Example Setup Configuration
Parameter
Setting
Select Supply
AMI 4Q06125PS
Stability Mode
Auto
Coil Constant
a
(kG/A)
a. Also referred to as the Field-to-Current Ratio. Obtained by
dividing the rated field by the rated current if not explicitly
stated.
0.934
Inductance (H)
9.7
Energy Absorber Present
NO
PSwitch Installed
YES
PSwitch Current (mA)
41.0
PSwitch Heated Time (sec)
20
PSwitch Cooled Time (sec)
20
PSwitch P/S Ramp Rate (A/sec)
10
PSwitch Cooling Gain (%)
0.0%
Current Limit (A)
85.600
Enable Quench Detect
YES
Extern Rampdown Enabled
YES
Voltage Limit (V)
3.500
b
b. Value is the 2.1 V charge rate plus allowances for power lead
drop (typically about 10-20 milliohms) at the rated current. If a
Model 601 energy absorber(s) is present, add an additional 5 V
to the value.
Ramp Rate (A/sec)
0.2165
c
c. Value is obtained by dividing the magnet charging voltage (V) by
the magnet inductance (H).
Содержание 430
Страница 2: ......
Страница 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Страница 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Страница 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Страница 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Страница 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...