156
R
EV
9
A
MERICAN
M
AGNETICS
, I
NC
.
R
EMOTE
I
NTERFACE
R
EFERENCE
: C
OMMAND
R
EFERENCE
S
WITCH
S
ETUP
C
ONFIGURATION
C
OMMANDS
AND
Q
UERIES
The Switch Setup Configuration Commands and Queries provide read/
write access to the setup functions available for the Switch Setup
submenu (page 83).
•
CONFigure:PSwitch {0|
1
}
•
PSwitch:INSTalled?
“0” indicates that a persistent switch is not installed on the connected
superconducting magnet. “1” indicates that a persistent switch is
installed. The default value is “1”. If a persistent switch is installed, the
persistent switch heated current should be specified. Heating/cooling
times should also be specified for timer-based switch transitions.
•
CONFigure:PSwitch:CURRent <
current (mA)
>
Sets the persistent switch heater current in mA.
•
PSwitch:CURRent?
Returns the persistent switch heater current setting in mA.
•
PSwitch:AUTODetect?
Executes the auto-detection algorithm (refer to page 84) and returns the
appropriate persistent switch heater current in mA. Note that after this
value is returned, it can be entered into the Model 430 Programmer
using the
CONFigure:PSwitch:CURRent
command.
NOTE
This remote command blocks communication on all ports until
the auto-detection process completes or is interrupted.
•
CONFigure:PSwitch:TRANsition {
0
|1}
Sending a value of “0” selects the timer-based switch transition
detection. Sending “1” selects the magnet voltage-based transition
detection. Timer-based transition detection is the default.
•
PSwitch:TRANsition?
“0” return value indicates timer-based switch transitions. “1” indicates
magnet voltage-based transitions.
•
CONFigure:PSwitch:HeatTIME <
time (sec)
>
Sets the time required in seconds for the persistent switch to become
resistive after the persistent switch heater has been activated.
Содержание 430
Страница 2: ......
Страница 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Страница 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Страница 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Страница 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Страница 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...