CONFigure:RAMP:RATE:FIELD <
segment>,<rate (kG/s, kG/min,
T/s, T/min)>,<upper bound (kG, T)
>
Sets the ramp rate for the specified segment (values of 1 through the
defined number of ramp segments are valid) in units of kilogauss/second
or minute, or tesla/second or minute (per the selected field units and
ramp rate units), and defines the field upper bound for that segment in
kilogauss or tesla (see page 95 for details of the use of ramp segments).
This command requires that a coil constant be defined; otherwise, an
error is generated.
•
RAMP:RATE:FIELD:<
segment>
?
Returns the ramp rate setting for the specified segment (values of 1
through the defined number of ramp segments are valid) in units of
kilogauss/second or minute, or tesla/second or minute (per the selected
field units and ramp rate units) and the current upper bound for that
range in kilogauss or tesla (per the selected field units). This command
requires that a coil constant be defined; otherwise, an error is generated.
The two return values are separated by a comma. For example:
Sent:
RAMP:RATE:FIELD:1?
Returned:
0.0100,5.0000
M
EASUREMENT
C
OMMANDS
AND
Q
UERIES
Included are queries for collecting the present supply current/voltage,
magnet field, magnet current, and magnet voltage.
•
VOLTage:MAGnet?
Returns the magnet voltage in volts. Requires voltage taps to be
installed across the magnet terminals and connected to a Magnet Station
rear connector.
•
VOLTage:SUPPly?
Returns the power supply voltage commanded by the Model 430
Programmer in volts.
•
CURRent:MAGnet?
Returns the current flowing in the magnet in amperes, expressed as a
number with a variable number of significant digits past the decimal point
(dependent on operating range), such as 5.2320. If the magnet is in
persistent mode, the command returns the current that was flowing in
the magnet when persistent mode was last entered.
Содержание 430
Страница 2: ......
Страница 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Страница 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Страница 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Страница 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Страница 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...