the mode to “On Entry”. Sending “2” sets the mode to “Cont f(T)”. “0” is
the default value.
•
OPLimit:MODE?
Returns “0” indicating the Protection Mode for the Operational Limits
submenu is Off. Returns “1” indicating the mode is “On Entry”. Returns
“2” indicating the mode is “Cont f(T)”.
•
CONFigure:OPLimit:ICSLOPE <
value (A/K)
>
Sets the
Ic Slope
value in the Operational Limits submenu. The units are
amperes/Kelvin.
•
OPLimit:ICSLOPE?
Returns the present
Ic Slope
value from the Operational Limits submenu
in units of amperes/Kelvin.
•
CONFigure:OPLimit:ICOFFSET <
value (A)
>
Sets the
Ic Offset
value in the Operational Limits submenu. The units are
amperes.
•
OPLimit:ICOFFSET?
Returns the present
Ic Offset
value from the Operational Limits submenu
in units of amperes.
•
CONFigure:OPLimit:TMAX <
value (K)
>
Sets the
T
max
value in the Operational Limits submenu. The units are
Kelvin.
•
OPLimit:TMAX?
Returns the present
T
max
value from the Operational Limits submenu in
units of Kelvin.
•
CONFigure:OPLimit:TSCALE <
value (K/V)
>
Sets the scale value in Kelvin/volts for the voltage-to-temperature
conversion for the voltage value read from Auxiliary Input 3 (see
page 202).
•
OPLimit:TSCALE?
Returns the scale value in Kelvin/volts for the Auxiliary Input 3 voltage-
to-temperature conversion.
Содержание 430
Страница 2: ......
Страница 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Страница 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Страница 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Страница 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Страница 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...