R
EV
9
113
A
MERICAN
M
AGNETICS
, I
NC
.
O
PERATION
: P
ERSISTENT
S
WITCH
C
ONTROL
P
ERSISTENT
S
WITCH
C
ONTROL
The Model 430 Programmer includes an integral persistent switch heater
that provides the capability of controlling the persistent mode of the
magnet either locally from the front panel of the Model 430 Programmer
using the
PERSIST. SWITCH CONTROL
key
46
, or remotely through a
communications interface. The persistent mode of the magnet is
indicated by the
MAGNET IN PERSISTENT MODE
LED
47
.
See page 58 for details of the use of the
PERSIST. SWITCH CONTROL
key.
NOTE
The following procedures include several automatically
sequenced steps once initiated. It is possible to interrupt the
process at any point by pressing the
SHIFT
+
ESC
keys. Please
note that the state in which the system will remain after this
interruption depends on the specific active step during which the
SHIFT
+
ESC
was pressed. Use this abort override keystroke with
caution.
P
ROCEDURE
FOR
I
NITIAL
H
EATING
OF
THE
S
WITCH
The Model 430 Programmer remembers the state of the persistent
switch during the time that the Programmer is de-energized. If the Model
430 is turned on when its shut down state was such that the persistent
switch was heated and commanding zero current (the normal state after
the magnet is discharged), the following screen will be displayed.
In order to charge the magnet, the persistent switch heater must be
energized. Perform the following steps.
46. See page 58.
47. See page 70.
0.00 A
P
Mode: Paused
0.00 Vs
PSwitch Heater: OFF
Содержание 430
Страница 2: ......
Страница 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Страница 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Страница 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Страница 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Страница 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...