8 Settings
- 81
–
3. Tap on
Measuring channels.
4. Untick the checkbox in the line of the measuring channel you want to
disable.
➢
The measuring channel is disabled and the measurement data of this
measuring channel will not be stored in the wireless sensor and not be
transferred to the data logger. The measuring channel will not be listed
in the menu Display.
Enable a measuring channel
Only measuring channels that have been preset but are currently
disabled can be enabled. In the wireless ALMEMO
®
sensor,
measuring channels are preset by default, in the wireless ALMEMO
®
interface precisely those measuring channels can be enabled that
are stored to the ALMEMO
®
D7 or D6 plug connected to the wireless
ALMEMO
®
interface.
Further measuring channels in the ALMEMO
®
D7 or D6 plug can be
enabled in the ALMEMO
®
Control software via a suitable ALMEMO
®
measuring device or via a ALMEMO
®
USB data cable ZA 1919
AKUV.
!
The wireless sensor needs to be connected to the data logger.
!
No measurement may be started.
1. In the home screen, tap on Sensor overview
.
2. Tap on the arrow at the end of the line of the wireless sensor, whose
measuring channel you want to enable.
3. Tap on
Measuring channels.
4. Tick the checkbox in the line of the measuring channel you want to
enable.
➢
The measuring channel is now enabled and the measurement data of
this measuring channel will be stored to the wireless sensor and then
be transferred to the data logger. The measuring channel is listed on
the menu Display.