Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
The RPG Knob is Not Functioning (N5181A/82A/83A)
1-61
Figure 1-29 A3J6 and A3J10
If the RPG test fails, complete the following procedure to determine the faulty assembly/part:
1.
Remove the front panel. Refer to
Chapter 3, "Assembly Replacement"
2.
Refer to
. Verify that A6J2-25 has a voltage of +3 Vdc:
•
If A6J2-25 has a voltage of +3 Vdc, go to
•
If A6J2-25 does not have a voltage of +3 Vdc, go to
.
3.
Refer to
. Measure A6J1-11 and A6J1-12; they should have a voltage of +3 Vdc.
•
If A6J1-11 and A6J1-12 have a voltage of +3 Vdc, replace the A6 DC-AC Inverter Interface Board.
•
If A6J1-11 and A6J1-12 do not have a voltage of +3 Vdc, go to
.
4.
Refer to
. Measure A3J6-11 and A3J6-12; they should have a voltage of +3 Vdc.
•
If A3J6-11 and A3J6-12 have a voltage of +3 Vdc, replace the W1 ribbon cable.
•
If A3J6-11 and A3J6-12 do not have a voltage of +3 Vdc, go to
.
5.
Refer to
. Measure A3J10-A21 and A3J10-A22; they should have a voltage of +3 Vdc.
•
If A3J10-A21 and A3J10-A22 have a voltage of +3 Vdc:
a.
Replace the A3 RF assembly.
b.
After replacing the A3 RF assembly in a model N5183A instrument, go to
Chapter 4, "Post-Repair Procedures and Performance
•
If A3J10-A21 and A3J10-A22 do not have a voltage of +3 Vdc, go to
“A1 Power Supply” on page 1-26
to troubleshoot the A1
Power Supply.
6.
Refer to
. Measure A6J2-28; it should have a voltage of +5 Vdc.
•
If A6J2-28 has a voltage of +5 Vdc, go to
•
If A6J2-28 does not have a voltage of +5 Vdc, go to
.
A3J611 and A3J6-12
A3J10-A21 and A3J10-A22
Front Panel
Rear Panel
Rear Panel
Front Panel
A3J6-3 and A3J6-4
A3J10-B16