Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
Inner Bottom Cover
3-8
Inner Bottom Cover
Tools Required
•
T-10 driver
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the outer-cover from the signal generator. Refer to
3.
Place the signal generator flat with the front-panel facing you.
4.
Using the T-10 driver, remove the eleven flat-top screws (1) from the inner bottom cover (2).
5.
Using the T-10 driver, remove the screw (3) from the rear panel.
6.
Remove the inner bottom cover.
Replacement Procedure
1.
Reverse the order of the removal procedures.
2.
Torque all screws to 9 in–lbs.