Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
No Output Power or Output Power is Unleveled (N5183A)
1-46
Low Output Power or No Output Power >= 3.2 GHz (N5183A)
1.
Remove the external cover and the top inside cover. Refer to
Chapter 3, "Assembly Replacement"
for instructions on removing the
instrument covers.
2.
, remove the W13 semirigid cable between the A3 RF output assembly and the 20 GHz doubler input located on the
A7 Micro Deck.
Figure 1-18 W13 Semirigid Cable
3.
Connect the A3 RF assembly output (see
) through a 10 dB attenuator to a spectrum analyzer.
4.
Set up the spectrum analyzer:
a.
Press
Preset
.
b.
Set the reference level to +10 dBm.
c.
Set the start frequency to 2.3 GHz.
d.
Set the stop frequency to 5.3 GHz.
e.
Set Trace to Max Hold.
5.
Set up the signal generator:
a.
Press
Preset
.
b.
Press
SWEEP
>
Sweep Type List Step
to Step.
c.
Press
Configure Step Sweep
.
•
Press
Freq Start
>
2.8
>
GHz
.
•
Press
Freq Stop
>
10.4
>
GHz
.
•
Press
# Points
>
501
>
Enter.
•
Press
More
>
Step Dwell
>
2
>
msec
.
•
Press
Return
>
Return
>
Sweep
>
Freq
Off
On
to On.
d.
Press
RF On/Off
to On.
e.
Press
AMPTD
> <maximum settable power> >
dBm
.
f.
Press
AMPTD
>
ALC Off On
to Off.
W13 semirigid cable
A3 RF assembly output