Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
A6 DC-AC Inverter Interface Board (N5181A/82A/83A only)
3-16
A6 DC-AC Inverter Interface Board (N5181A/82A/83A only)
Tools Required
•
T-10 driver
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the covers from the signal generator. Refer to
.
3.
Remove the front-panel from the signal generator. Refer to
.
4.
Using the T-10 driver, remove the four screws (1) that secure the inverter shield (2) to the A6 DC-AC Inverter Interface Board (3).
5.
Disconnect the W1 ribbon cable (not shown in
) from the A6 DC-AC Inverter Interface board (3).
6.
Disconnect the Power Switch cable (4) from the A6 DC-AC Inverter Interface Board (3).
7.
Disconnect the LCD 2 wire cable (5) from the A6 DC-AC Inverter Interface Board (3).
8.
Disconnect the ribbon cable (6) from the A6 DC-AC Inverter Interface Board (3).
9.
Disconnect the LCD flat flex cable (7) from the A6 DC-AC Inverter Interface Board (3).
10. Using the T-10 driver, remove the three screws (8) that secure the A6 DC-AC Inverter Interface Board (3) to the front-panel sub-panel
(9).
11. Remove the A6 DC-AC Inverter Interface Board.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Torque all screws to 9 in–lbs.
3.
Perform the post-repair performance tests that pertain to this replacement procedure, see
Chapter 4, Post-Repair Procedures and
.
135-1095-0799