Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
No Output Power or Output Power is Unleveled (N5183A)
1-40
Figure 1-15 A3J6-3 and A3J10-B16
10. Refer to
. Using the voltmeter, measure A3J10-B16.
•
If A3J10-B16 is
∼
12V, perform the following steps:
a.
Replace the A3 RF assembly.
b.
Chapter 4, "Post-Repair Procedures and Performance Verification"
•
If A3J10-B16 is not
∼
12V, replace the A1 Power Supply.
No Output Power or Output Power is Unleveled (N5183A)
1.
Connect the signal generator’s RF output through a 10 dB attenuator to a spectrum analyzer.
2.
Set up the spectrum analyzer:
a.
Press
Preset
.
b.
Set the reference level to +25 dBm.
c.
Set the start frequency to 50 kHz.
d.
Set the stop frequency to the signal generator’s maximum frequency.
e.
Set trace to Max Hold.
3.
Set up the signal generator:
a.
Press
Preset
.
b.
Press
SWEEP
>
Sweep Type List Step
to Step.
c.
Press
Configure Step Sweep
.
•
Press
Freq Start
>
100
>
kHz
.
•
Press
Freq Stop
> <maximum_signal_generator_frequency> >
GHz.
•
Press
# Points
>
501
>
Enter.
•
Press
More
>
Step Dwell
>
2
>
msec
.
A3J6-4
A3J10-B16
Front Panel
Rear Panel
Rear Panel
Front Panel
深圳德标仪器