Agilent Technologies N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Post-Repair Procedures and Performance Verification
Leveled Output Power Verification (N5161A/62A
1
/81A/82A Only)
4-14
7.
Set the amplitude to 13
1
dBm:
Press
Amplitude
>
13
>
dBm
.
8.
Configure the power meter for the measurement.
a.
Press the
Frequency Cal Fac
button on the power meter.
b.
Select a power meter channel (if applicable).
c.
Use the arrow keys to enter the frequency at which to measure the power.
9.
Measure the output power level.
10. Repeat steps
to measure power at each of the 15 frequencies listed in
11. Confirm that the measured power levels are within the limits listed in the table.
NOTE
The limit range is due to power meter uncertainty.
538 MHz
<
0.6dB
675 MHz
<
0.6dB
850 MHz
<
0.6dB
1075 MHz
<
0.6dB
1350 MHz
<
0.6dB
1700 MHz
<
0.6dB
2150 MHz
<
0.6dB
2700 MHz
<
0.6dB
3400 MHz
<
0.7dB
4300 MHz
<
0.8dB
5400 MHz
<
0.8dB
1
For serial numbers <MY4818xxxx, US4818xxxx, and SG4818xxxx, the amplitude = 7 dBm
.
Table 4-10Leveled Output Power Limits (Continued)
Frequency
Limits (dB)