84
1260 SFC Binary Pump
6
Error Information
Module Error Messages
Encoder Missing
Error ID: 2046, 2050, 2510
Encoder Missing:
Pump channel A
B: Encoder Missing:
Pump channel B
The optical encoder on the pump motor in the module is missing or defective.
The processor checks the presence of the pump encoder connector every 2 s. If
the connector is not detected by the processor, the error message is generated.
Probable cause
Suggested actions
1
Defective or disconnected pump encoder
connector.
Please contact your Agilent service
representative.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.