1260 SFC Binary Pump
119
Maintenance
8
Exchange the Inlet Valve
Exchange the Inlet Valve
When
If internally leaking (backflow)
Tools required
Description
Wrench, 14 mm
Parts required
p/n
Description
G1312-60066
Passive inlet valve
G4302-60066
Inlet Valve SFC
Preparations
Switch off the pump at the main power switch
1
Disconnect the solvent inlet tube at the inlet valve
(beware of leaking solvents).
2
Using a 14 mm wrench, loosen the active inlet valve and
remove the valve from the pump head.
3
Screw the new valve into the pump head and tighten
using the 14 mm wrench. Do not overtighten.
4
Reconnect the inlet tube to the valve.