1260 SFC Binary Pump
87
Error Information
6
Module Error Messages
Index Adjustment
Error ID: 2204, 2214
Index Adjustment:
Pump channel A
B: Index Adjustment:
Pump channel B
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the time to reach the index position is too long,
the error message is generated.
Index Missing
Error ID: 2205, 2215, 2505
Index Missing:
Pump channel A
B: Index Missing:
Pump channel B
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is not recognized within a
defined time, the error message is generated.
Probable cause
Suggested actions
1
Irregular or sticking drive movement.
Remove the pump head, and examine the
seals, pistons, and internal components for
signs of wear, contamination or damage.
Exchange components as required.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Disconnected or defective encoder cable.
Please contact your Agilent service
representative.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...