100
1260 SFC Binary Pump
8
Maintenance
Exchange the Purge Valve Frit or the Purge Valve
Exchange the Purge Valve Frit or the Purge Valve
When
•
Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar in low delay volume configuration and > 20 bar in standard configuration across the
frit at a flow rate of 5 mL/min of water with purge valve opened)
•
Purge valve – if internally leaking
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-1924
Wrench open 14 mm
Pair of tweezers
OR
Toothpick
Parts required
p/n
Description
Frits
Purge Valve
5067-4728
Seal cap
Preparations
•
Switch off pump at the main power switch
•
Open the doors
•
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages
1
Using a 1/4 inch wrench disconnect the pump outlet
capillary from the purge valve. Disconnect the waste
tube. Beware of leaking solvents due to hydrostatic
pressure.
2
Using the 14 mm wrench, unscrew the purge valve and
remove it from the purge valve holder.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...