1260 SFC Binary Pump
3
Using the Pump
Setting up the Pump with the Instrument Control Interface
For details, please refer to the online help or manual of your instrument
control software.
Bottle Filling
The pump offers a powerful feature to monitor the liquid level in the solvent
bottles. With total bottle volume and initial filling volume set correctly, the
pump subtracts the consumed volume continuously from the initial value and
stops the pump and method/sequence execution before the system runs dry or
an analysis is corrupted.
on page 44 lists the available bottle filling parameters.
N O T E
The pressure data curve is
generated
from the pressure sensor readings, while %A, %B
and flow are
calculated
from the method settings of the pump.
C A U T I O N
The bottle filling feature fails if multiple solvent inlets are put into one solvent bottle!
➔
In that case implement a minimum pressure limit to avoid that the pump runs dry
when solvents are empty.
Table 5
Bottle Filling Parameters
Parameter
Limits
Description
•
Total
Volume
0 – 1000 L
default: 0 L
This is the capacity (maximum possible volume) in liter of the solvent bottle. In
combination with the actual volume, this parameter is used for calculating and
displaying the relative liquid level.
•
Actual
Volume
0 – 1000 L
default: 0 L
After filling the solvent bottles, enter the actual volumes into these boxes. The
Actual Volume
must not be larger than the
Total Volume
of the bottle.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...