36
1260 SFC Binary Pump
3
Using the Pump
Best Practices
Prepare the pump
Purge
Use the Purge function to:
•
fill the pump,
•
exchange a solvent,
•
remove air bubbles in tubes and pump heads.
Seal wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
•
When using buffers with elevated salt concentrations
•
When using volatile solvents with non-volatile additives
How to deal with solvents
•
Use clean bottles only.
•
Exchange water-based solvents daily.
•
Select solvent volume to be used up within 1 – 2 days.
•
Use only HPLC-grade solvents and water filtered through 0.2 μm filters.
•
Label bottles correctly with bottle content, and filling date / expiry date.
•
Use solvent inlet filters.
•
Reduce risk of algae growth: use brown bottles for aqueous solvents, avoid
direct sunlight.
C A U T I O N
Contaminated seal wash solvent
➔
Do not recycle seal wash solvent to avoid contamination.
➔
Weekly exchange seal wash solvent.