1260 SFC Binary Pump
181
LAN Configuration
12
Dynamic Host Configuration Protocol (DHCP)
2
Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Card or the main board of above mentioned modules.
3
Turn on the module that hosts the LAN interface.
4
Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition,
Lab Advisor, Firmware Update Tool) and use MAC address as host name,
e.g.
0030d3177321
.
The LC system should become visible in the control software (see Note in
section
Table 18
G1369C LAN Interface Card (configuration switch on the card)
SW 4
SW 5
SW 6
SW 7
SW 8
Initialization Mode
ON
OFF
OFF
OFF
OFF
DHCP
Table 19
LC Modules with 8-bit configuration switch (B-firmware) (configuration switch
at rear of the instrument)
SW 6
SW 7
SW 8
Initialization Mode
ON
OFF
OFF
DHCP