114
1260 SFC Binary Pump
8
Maintenance
Reinstall the Pump Head Assembly
Reinstall the Pump Head Assembly
When
When reassembling the pump
Tools required
•
3-mm hexagonal key
•
4-mm hexagonal key
1
Slide the pump head assembly onto the pump drive.
2
Using a 4 mm hexagonal key, tighten the pump head
screws stepwise with increasing torque.
3
Reconnect all tubings and capillaries.
4
Close the doors.