78
1260 SFC Binary Pump
6
Error Information
Module Error Messages
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
Pressure Signal Missing
Error ID: 2016
The pressure signal of the damper is missing.
The pressure signal of the damper must be within a specific voltage range. If
the pressure signal is missing, the processor detects a voltage of
approximately -120 mV across the damper connector.
Probable cause
Suggested actions
1
Lower pressure limit set too high.
Ensure the lower pressure limit is set to a value
suitable for the analysis.
2
Air bubbles in the mobile phase.
•
Make sure that the degasser is in flow path
and works correctly. Purge the module.
•
Ensure solvent inlet filters are not blocked.
3
Leak.
•
Inspect the pump head, capillaries and
fittings for signs of a leak.
•
Purge the module. Run a pressure test to
determine whether the seals or other
module components are defective.
4
Defective damper.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Damper disconnected.
Please contact your Agilent service
representative.
2
Defective damper.
Please contact your Agilent service
representative.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...