1260 SFC Binary Pump
51
Using the Pump
3
Solvent Information
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents.
There are no documented incompatibilities for HPLC applications.
Fluorinated polymers (PTFE, PFA, FEP, FFKM)
Fluorinated polymers like PTFE (polytetrafluorethylene), PFA
(perfluoroalkoxy) and FEP (fluorinated ethylene propylene) are inert to
almost all common acids, bases, and solvents. FFKM is perfluorinated rubber,
which is also resistant to most chemicals. As an elastomer, it may swell in
some organic solvents like halogenated hydrocarbons.
TFE/PDD copolymer tubings, which are used in all Agilent degassers except
1322A, are not compatible with fluorinated solvents like Freon, Fluorinert, or
Vertrel. They have limited life time in the presence of Hexafluoroisopropanol
(HFIP). To ensure the longest possible life with HFIP, it is best to dedicate a
particular chamber to this solvent, not to switch solvents, and not to let dry
out the chamber. For optimizing the life of the pressure sensor, do not leave
HFIP in the chamber when the unit is off.
Sapphire, Ruby and Al
2
O
3
-based ceramics
Sapphire, ruby and ceramics based on aluminum oxide Al
2
O
3
are inert to
almost all common acids, bases and solvents. There are no documented
incompatibilities for HPLC applications.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...