166
1260 SFC Binary Pump
11
Hardware Information
Setting the 6-bit Configuration Switch
Setting the 6-bit Configuration Switch
The 6-bit configuration switch is located at the rear of the module with
FUSION electronics. Switch settings provide configuration parameters for
LAN and instrument specific initialization procedures.
All modules with FUSION electronics:
•
Default is ALL switches DOWN (best settings).
•
Default IP address for LAN 192.168.254.11
•
For specific LAN modes switches 4-5 must be set as required.
•
For boot resident/cold start modes switches 1+2 or 6 must be UP.
Figure 16
Location of Configuration switch (example shows a G7114A/B VWD)
&RQILJXUDWLRQVZLWFK