124
1260 SFC Binary Pump
8
Maintenance
Replacing Module Firmware
Replacing Module Firmware
To upgrade/downgrade the module’s firmware carry out the following
steps:
1
Download the required module firmware, the latest FW Update Tool and
the documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2
For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
There is no specific information for this module.
When
The installation of newer firmware might be necessary
•
if a newer version solves problems of older versions or
•
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
•
to keep all systems on the same (validated) revision or
•
if a new module with newer firmware is added to a system or
•
if third party control software requires a special version.
Tools required
Description
Agilent Lab Advisor software
OR
Instant Pilot G4208A
(only if supported by module)
Parts required
#
Description
1
Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...