94
Agilent E8462A Relay Multiplexer Command Reference
one channel per MUX can be closed at any time in the WIRE1 modes due to the need
of the control relay to choose the “side” for connection to the common terminal. For
example, in the WIRE1 mode (one 1X256), the control relay chooses either Ter0 or
Ter1 for connection to the one-wire common terminal (Ter0). You should not make
any field connection to the odd numbered common terminals (e.g., Ter1, 3, 5, 7, 9,
11, 13 or 15) because these connectors do get routed to the channels by the control
relay.
Channel numbers 0 to 99 in MUX 0:
Channel numbers from 0 to 99 in MUX 0 can
be referred to by ccnn. The long-hand version would be ccbnnn. For MUX 0, b = 0
and the first n = 0 for channels 0 to 99. The leading zeros in the bnnn specification
can be deleted in pairs to shorten the channel specification.
Opening Channels:
•
Any combination of the above channel lists is valid in one command.
•
Opening order for multiple channels with a single command is not guaranteed.
Note
Channel numbers can be in the
channel_list
in any random order but if the
sequential channel list is used (ccbnnn:ccbnnn), the second channel listed must be
greater than the first channel specified.
Opening the Control Relays:
The control relays (9000 to 9021, 9100 to 9108, 9200
to 9204) can be opened to perform special functions (for example, isolating channels
from the analog bus). However, if the multiplexer has not been configured to the
NONE mode with
[ROUTe:]FUNCtion
command, then doing a CLOSe or SCAN of
any bank switch channel will automatically close the associated control relays.
Open:
9000 to 9021
to disconnect the associated bank of relays from the common terminals
(Ter0 - Ter15). These are the T0 to T21 switches of Figure 1-1.
9100 to 9107
to disconnect the low side of the banks from the high terminal. These
are the C100 to C107 form-C switches of Figure 1-1. These switches are used for
making one-wire connections.
9108
to disconnect the low side of the banks from the high terminal when making
one-wire measurements and the scan port is set to the analog bus. This switch is
switch C108 in Figure 1-1.
9200 to 9204
to disconnect the banks from the analog bus. These switches are
labeled AB200 - AB204 in Figure 1-1.
- 9200
disconnects analog H1 from switch C108 (9108).
To Open:
Use this command:
a single channel
[ROUT:]OPEN (@ccbnnn)
multiple channels
[ROUT:]OPEN (@ccbnnn,ccbnnn,...)
sequential channels
[ROUT:]OPEN (@ccbnnn:ccbnnn)
groups of sequential
channels
[ROUT:]OPEN (@ccbnnn:ccbnnn, ccbnnn:ccbnnn)
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...