90
Agilent E8462A Relay Multiplexer Command Reference
20067 closes channel 67 of card #2.
[ROUTe:]CLOSe?
[ROUTe:]CLOSe? <
channel_list
>
Returns the current state of the channel(s)
queried.
Channel_list
has the form (@ccbnnn) (see
[ROUTe:]CLOSe
for definition).
The command returns "1" if channel(s) are closed or returns "0" if channel(s) are
open.
Comments
Query is Firmware Readback:
The
ROUTe:CLOSe?
command returns the current
firmware state of the channel(s) specified. It does not account for relay hardware
failures. A maximum of 128 channels at a time can be queried for any switchbox.
To query all the channels in the WIRE1 mode requires at least two queries (e.g.,
10000:10127 and 10128:10255).
Example
Query Channel Closure
This example closes channel 0 in card 01, and channel 67 in card 02 of a two-module
switchbox and queries channel closure. Since the channels are programmed to be
closed "1,1" is returned.
CLOS (@10000,20067)
10000 closes channel 0, card #1, and
20067 closes channel 67, card #2.
CLOS? (@10000,20067)
Query state of channel 0, card #1; and
channel 67, card #2.
[ROUTe:]FUNCtion
[ROUTe:]FUNCtion <
card_number
>, <
function
>
Selects the operating mode of the
multiplexer channels. All channels on the card specified by
card_number
operate in
the specified mode.
ROUTe:
is NOT optional when
ROUT:FUNC
is used with a
scanning multimeter configuration.
Parameters
Comments
Defaults to WIRE2 mode:
This command is required if you want to set a
configuration other than 2-wire.
ROUTe:
is Not Always Optional:
If used with a scanning multimeter configuration,
ROUTe:FUNCtion
must be used.
Using the
FUNCtion
Command:
Using the
FUNCtion
command to reconfigure the
multiplexer, the command must be sent
EACH TIME
the card is powered up.
Parameter
Name
Parameter
Type
Range of Values
Default
Value
card_number
numeric
01 to 99
N/A
function
discrete
WIRE1|WIRE2|WIRE3|WIRE4|
WIRE1X2|WIRE2X2|WIRE4X2|
WIRE1X4|WIRE2X4|WIRE4X4
WIRE1X8|WIRE2X8|NONE
WIRE2
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...