Using the Multiplexer
Recalling and Saving States
This section contains information about saving and recalling a Multiplexer
module state.
Saving States
The
*SAV
<numeric_state>
command saves the current instrument state.
The state number (0-9) is specified in the state parameter. The following
settings are saved:
•
Channel Relay State (CH000-256 open or closed);
•
Tree Relay State (CH9000-9021 open or closed);
•
Routing Relay Stare (CH9100-9108 open or closed).
•
Analog Bus Connection Control Relay State (CH9200-9204 open or
closed)
•
ARM:COUNt Value
•
TRIGger:SOURce Mode
•
OUTPut[:STATe] Configuration
•
INITiate:CONTinuous Mode
•
[ROUTe:]FUNCtion Mode
•
[ROUTe:]SCAN:MODE Mode
•
[ROUTe:]SCAN:PORT Mode
•
DIAGnostic:SCAN:DELay Time
•
DIAGnostic:INTerrupt[:LINE] Setting
•
DIAGnostic:INTerrupt:TIME Time
Recalling States
The
*RCL
<numeric_state>
command recalls a previously saved state.
Enter the number (0-9) in the state parameter of the desired saved state. If
*SAV
was not previously executed using the selected number, the
Multiplexer will be configured to the reset values (see "Table 2-1. Agilent
E8462A Default Conditions for Power-on and Reset" on page 35).
Note
Scan lists are not saved when a state is saved. You must re-enter your scan
list after recalling a state.
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...