98
Agilent E8462A Relay Multiplexer Command Reference
[ROUTe:]SCAN:MODE
[ROUTe:]SCAN:MODE <
mode
>
Sets the multiplexer channels defined by the
[ROUTe:]SCAN <
channel_list
>
command for none, volts, two-wire ohms, or
four-wire ohms measurements.
Parameters
Comments
Order of Command Execution:
The
[ROUTe:]SCAN:MODE
and
[ROUTe:]FUNCtion
commands must be executed before the
[ROUTe:]SCAN
<
channel_list
>
command.
RES Mode versus 4-wire
:
When
[ROUTe:]FUNC
is WIRE4, WIRE4X2 OR
WIRE4X4, RES Mode will operate the same as FRES.
NONE
and
VOLT
Mode:
When selected,
channel_list
is setup for volts
measurements.
VOLT
mode is also used when making two-wire ohms measurements
using two-wire multimeters.
RES
Mode in WIRE2:
When selected,
channel_list
is setup for two-wire ohms
measurements. Control relays 9002 and 9009 are closed to connect Terminal 4
(Ter4) to Terminal 0 (Ter0) and Terminal 5 (Ter5) to Terminal 1 (Ter1). Ter4 and
Ter5 could be used to connect to the Ohmmeter’s current source. When
SCAN:PORT ABUS
is selected, control relays 9203 and 9204 are also closed to
connect Analog H2 and Analog L2 which are the Analog Bus connections to the
Ohmmeter’s current source. When selected, the multimeter SENSE/SOURCE leads
are used to make the measurement. When using the HI/LO leads on a multimeter to
make the measurement, use the
VOLT
mode.
FRES
Mode:
When selected,
channel_list
is setup for four-wire ohms
measurements. This mode is supported by the WIRE4, WIRE4X2 and WIRE4X4
FUNCtions. All channels of WIRE4 may be routed to the Analog Bus, but only the
lowest bank of WIRE4X2 and WIRE4X4 may be routed to the Analog Bus.
*RST Condition:
[ROUTe:]SCAN:MODE NONE
Example
Selecting the Four-Wire Ohms Measurement
This example selects the four-wire ohms measurement mode (
FRES
) on card #1 of
a single-module switchbox.
FUNC 1,WIRE4
Set mode to four-wire.
TRIG:SOUR EXT
Selects external trigger source.
SCAN:MODE FRES
Selects four-wire
scan mode.
SCAN (@130:137)
Scan channels 30 to 37.
INIT
Starts scanning cycle.
Parameter
Name
Parameter
Type
Range of Values
Default
Value
mode
discrete
NONE|VOLT|RES|FRES
NONE
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...