Agilent E8462A Relay Multiplexer Command Reference
*RST Condition:
OUTPut:ECLTrg
n
[:STATe]
OFF
(disabled)
Example
Enabling ECL Trigger Bus Line 0
OUTP:ECLT0:STAT 1
Enable ECL Trigger bus line 0 to output
pulse after each scanned channel is
closed.
OUTPut:ECLTrgn[:STATe]?
OUTPut:ECLTrg
n
[:STATe]?
Queries the present state of the specified ECL Trigger
bus line. The command returns "1" if the specified bus line is enabled or "0" if the
specified bus line is disabled.
Example
Query ECL Trigger Bus Enable State
This example enables ECL Trigger bus line 0 and queries the enable state. The
OUTPut:ECLTrg
n
?
command returns "1" since the port is enabled.
OUTP:ECLT0:STAT 1
Enable ECL Trigger bus line 0.
OUTP:ECLT0?
Query bus enable state.
OUTPut[:EXTernal][:STATe]
OUTPut[:EXTernal][:STATe] <
mode
>
Enables or disables the "Trig Out" port on
the Agilent E1406A Command Module to output a trigger when a channel is closed
during a scan.
ON|1
enables the port and
OFF|0
disables the port.
Parameters
Comments
Enabling "Trig Out" Port:
When enabled, a pulse is output from the "Trig Out"
port after each scanned switchbox channel is closed. If disabled, a pulse is not output
from the port after channel closures. The output is a negative-going pulse.
"Trig Out" Port Shared by Switchboxes:
Only one switchbox configuration can
use the selected trigger at a time. When enabled, the "Trig Out" port is pulsed by the
switchbox each time a scanned channel is closed. To disable the output for a specific
switchbox, send the
OUTP OFF
or
0
command for that switchbox.
One Output Selected at a Time:
Only one output (
ECLTrg0
or
1
;
TTLTrg0
,
1
,
2
,
3
,
4
,
5
,
6
, or
7
; or
EXTernal
) can be enabled at one time. Enabling a different output
source will automatically disable the active output. For example, if
TTLTrg1
is the
active output and
TTLTrg4
is enabled,
TTLTrg1
will become disabled and
TTLTrg4
will become the active output.
Related Commands:
[ROUTe:]SCAN
,
TRIGger:SOURce
,
OUTPut[:EXTernal][:STATe]?
Parameter
Name
Parameter
Type
Range of Values
Default
Value
mode
boolean
0|1|OFF|ON
OFF|0
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...