Agilent E8462A Relay Multiplexer Command Reference
87
[ROUTe:]
The
[ROUTe:]
command subsystem controls switching and scanning operations for
multiplexer modules in a switchbox.
Note
This command opens all previously closed relays, therefore, it should be the first
relay configuration command.
Subsystem Syntax
[ROUTe:]
CLOSe <
channel_list
>
CLOSe? <
channel_list
>
FUNCtion <
card_number
>, <
function
>
FUNCtion? <
card_number
>
OPEN <
channel_list
>
OPEN? <
channel_list
>
SCAN <
channel_list
>
:MODE <
mode
>
:MODE?
:PORT <
port
>
:PORT?
[ROUTe:]CLOSe
[ROUTe:]CLOSe <
channel_list
>
Closes the multiplexer channels specified by
channel_list
.
Channel_list
has the form (@ccbnnn) where cc = card number (01-99),
b = bank or MUX number (0 - one less than number of muxes), and nnn = channel
number (0 - one less than number of switches per MUX).
Parameters
Parameter
Name
Parameter
Type
Mode
Range of Values
channel_list
numeric
WIRE1
WIRE2
WIRE3
WIRE4
cc0000 to cc0255
cc0000 to cc0127
cc0000 to cc0063
cc0000 to cc0063
WIRE1X2
cc0000 to cc0127
cc1000 to cc1127
WIRE2X2
cc0000 to cc0063
cc1000 to cc1063
WIRE4X2
cc0000 to cc0031
cc1000 to cc1031
WIRE1X4
cc0000 to cc0063
cc1000 to cc1063
cc2000 to cc2063
cc3000 to cc3063
table is continued on next page
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...