Agilent E8462A Specifications
125
Appendix A
Agilent E8462A Specifications
General Characteristics
Module Size/Device Type
C-Size VXIbus, Register based, A16/D16
Interrupt Level
1-7, selectable
Cooling/Slot
Watts/slot:
30W max, 10W typical
P mm H
2
O:
0.1
Air Flow (liters/sec): 3.0
Operating Temperature
0 - 55°C
Operating Humidity
65% RH, 0 - 40°C
Operating Location
Intended for indoor use only. Operating location should be a sheltered
location where air temperature and humidity are controlled within this
product’s specifications and the product is protected against direct exposure
to climatic conditions such as direct sunlight, wind, rain, snow, sleet and
icing, water spray or splash, hoarfrost (frost) or dew.
Pollution Environment
Pollution environment for which this product may be operated is IEC 664
Pollution Degree 2 (typically, indoor). Pollution degree 2 means only
non-conductive pollution occurs. However, occasionally a temporary
concuctivity caused by condensation must be expected.
Power Requirements
Voltage
I
pm
(A)
I
dm
(A)
+5V
1.0
1
0.50
+12V
0.0
0.0
-12V
0.0
0.0
Relay Life
(typical)
NOTE:
Relays are subject to normal wear out based on the number of operations.
1. Specified in the worst case typical mux configuration (eight 32 x 1). Add 30 mA per additional relay.
Condition
Number of
Operations
1.0 V & 10 mA
10 x 10
5
Rated Full Resistive Load (60VA)
10 x 10
4
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...