100
Agilent E8462A Relay Multiplexer Command Reference
[ROUTe:]SCAN:PORT ABUS
with 2-Wire Functions:
With FUNC set to
WIRE2, WIRE2X2, WIRE2X4 or WIRE2X8 and SCAN:PORT set to ABUS,
control relays 9200 and 9201 are closed. In addition, if the SCAN:MODE is not
RES, then control relays 9203 and 9204 are closed. If SCAN:MODE is not RES,
then Analog Control Relay 9202 is closed. Analog Control Relays that are not listed
as being closed will be opened.
[ROUTe:]SCAN:PORT ABUS
with 3-Wire Functions:
With FUNC set to
WIRE3 and SCAN:PORT set to ABUS, control relays 9200, 9201 and 9204 are
closed. Control relays 9202 and 9203 are opened.
[ROUTe:]SCAN:PORT ABUS
with 4-Wire Functions:
With FUNC set to
WIRE4, WIRE4X2, or WIRE4X8 and SCAN:PORT set to ABUS, control relays
9200, 9201, and 9204 are closed. Control relay 9202 is opened. If SCAN:MODE is
VOLT, then Analog Control Relay 9203 is opened; if the mode is not VOLT, 9203
is closed.
*RST Condition:
[ROUTe:]SCAN:PORT NONE
Example
Selecting the Analog Bus Port
This example selects the four-wire ohms measurement mode (
FRES
) on card #1 of
a single-module switchbox, then enables the analog bus connection. Control relays
9200, 9201, 9203 and 9204 will close, and 9202 will open.
FUNC 1,WIRE4
Set mode to four-wire.
TRIG:SOUR EXT
Selects external trigger source.
SCAN:MODE FRES
Selects the four-wire ohms mode.
SCAN:PORT ABUS
Selects the analog bus port.
SCAN (@130:137)
Scan channels 30-37.
INIT
Start scanning cycle.
[ROUTe:]SCAN:PORT?
[ROUTe:]SCAN:PORT?
Returns the current state of the analog bus port. The
command returns
NONE
if the analog bus connection control relays are disabled or
ABUS
if the control relays are enabled.
Example
Query the Scan Port
This example selects the analog bus port, then queries the state. Because the analog
bus port is selected, the query command returns "ABUS".
SCAN:PORT ABUS
Selects the analog bus port.
SCAN:PORT?
Query the port selection.
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...