18
Configuring the Agilent E8462A Multiplexer
Connecting Field Wiring
Field Wiring
Not Using Optional
Terminal Cards
The E8462A
IS NOT
supplied with a terminal card or connectors. You may
purchase 160-pin terminal connectors, necessary crimp-and-insert contacts
and the required crimp tool from Agilent Technologies or directly from the
manufacturer, ERNI Components
1
(see Table 1-1 below).
Field Wiring
Terminal Connector
Refer to Table 1-1 and Figure 1-5. You can purchase 160-pin field wiring
terminal connectors (two are required, order Agilent P/N 1252-6531 or order
direct from the manufacturer, ERNI Components, P/N 024070) and the
necessary crimp-and-insert contacts (Agilent single contact P/N is 1252-
6533, or ERNI P/N 014728). The contacts are gold-plated, accept a wire size
of 20 to 26AWG, and carry a maximum current of 2A @70°C. You will also
need a crimp tool (Agilent P/N 8710-2306 or ERNI P/N 014374) and option-
ally a disassembly tool (Agilent P/N 8710-2307 or ERNI P/N 471555).
Caution
Due to the close terminal spacing and the potential for pin-to-
pin leakage, the terminal connector blocks on the Option 012
Crimp-and-Insert Terminal Card must be replaced after 15,000
hours of voltage stress if the module regularly switches
voltages greater than 60VDC, 50VACrms, or 70.7 VACpeak.
Table 1-1. You Must Purchase Connectors, Contacts and Tools.
Manufacturer
Component
Connector P/N
Agilent Tecnologies
160-pin connector
1252-6531
one (1) crimp-and-
insert contact
1252-6533
one (1) crimp-and-
insert contact single
conductor assembly
8150-5207
(see Figure 1-5)
required crimp tool
8710-2306
disassembly tool
(optional)
8710-2307
ERNI Components
1
160-pin connector
024070
one (1) crimp-and-
insert contact
014728
required crimp tool
014374
disassembly tool
(optional)
471555
1. Contact ERNI Components, A Division of ODIN Components, Inc., 520 Southlake Blvd., Richmond, VA
23236, U.S.A. Telephone, (804) 794-6367, FAX (804) 379-2109.
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...