16
Configuring the Agilent E8462A Multiplexer
Setting the Interrupt
Priority Line
For most applications the default priority line should not have to be changed.
An interrupt is generated after any channel is opened or closed when
interrupts are enabled. The interrupt is generated approximately 5 ms after
command execution allowing for relay settling time. The interrupt line can
be set to any one of the VXI backplane lines 1-7 through writing the bits 10,
9 and 8 of the Status/Control Register. The default value is 1. The interrupt
can be disabled at power-up, after a SYSRESET, or after resetting the
module via the Control Register.
See the
DIAGnostic:INTerrupt[:LINE]
command for setting the interrupt
priority line using a SCPI command. See
Appendix B, Agilent E8462A
Register-Based Programming
, for more information about setting the
interrupt priority line by writing to the Status/Control Register.
Protection
Resistors
Figure 1-1 shows the 100
protection resistors in series with each bank line.
These protection resistors limit the maximum current through the relays.
However, in some measurements (such as 2-Wire resistance measurements)
you may want to bypass the protection resistors. Each resistor has a jumper
(J601 for Bank 0, JP602 for Bank 1, JP 603 for Bank2, . . . JP616 for Bank
15) across it allowing you to short out the resistor if necessary. Refer to
Figure 1-3.
Installing Protection
Resistor Jumpers
The module is shipped from the factory with a bag of jumpers. You must
install these on the E8462A PC board in one of the two settings shown in
Figure 1-3. It is recommended you install them in the default position shown
in Figure 1-3 if your application requires 100
protection resistors. Install
them over both jumper pins if you want to bypass the 100
protection
resistors.
Note
The module is shipped from the factory with a bag of jumpers. You should
load these jumpers in the default position as shown in Figure 1-3 if your
application requires 100
protection resistors. Install the jumpers across
both pins if you do not require the 100
protection resistors in your
application.
Figure 1-3. Protection Resistors and Jumpers
=
=
(default)
JUMPERS ARE SUPPLIED IN A BAG
AND MUST BE INSTALLED
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...