Agilent E8462A Relay Multiplexer Command Reference
75
DIAGnostic:INTerrupt[:LINE]
DIAGnostic:INTerrupt[:LINE]
<card_number>, <line_number>
sets Multiplexer
interrupt line. The
card_number
specifies which Agilent E8462A in a
multiple-module switchbox, is being referred to. The
line_number
can be 1 through
7 corresponding to VXI backplane interrupt line 1-7.
Parameter
Comments
•
Setting
<line_number>
= 0 will disable the Multiplexer’s interrupt.
•
Only one value (1 through 7) can be set at one time.
•
The default value of
<line_number>
is 1 (lowest interrupt line).
Example
Setting the Multiplexer’s interrupt line equal to interrupt line 6.
DIAG:INT:LINE 1, 6
Set the interrupt line equal to line 6.
DIAGnostic:INTerrupt[:LINE]?
DIAGnostic:INTerrupt[:LINE]? <card_number>
queries the module’s VXI
backplane interrupt line and the return value is one of 1, 2, 3, 4, 5, 6, 7 which
corresponding to the module’s interrupt line 1-7. The return value being 0 indicates
that the Multiplexer is interrupt disabled. The
card_number
specifies which Agilent
E8462A in a multiple-module switchbox, is being referred to.
Parameter
Comments
•
Return value of “0” indicates that the Multiplexer’s interrupt is disabled.
Return values of 1-7 correspond to VXI backplane interrupt lines 1 through 7.
•
When power-on or reset the module, the default interrupt line is 1.
Example
Query the Multiplexer’s interrupt line.
DIAG:INT:LINE 1, 6
Set the interrupt line equal to 6.
DIAG:INT:LINE?
Query the Multiplexer’s interrupt line.
Name
Type
Range of Values
Default Value
<card_number>
<line_number>
numeric
numeric
1 - 99
0 - 7
1
1
Name
Type
Range of Values
Default Value
<card_number>
numeric
1 - 99
1
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...