Agilent E8462A Relay Multiplexer Command Reference
85
*RST Condition:
OUTPut[:EXTernal][:STATe] OFF
(disabled)
Example
Enabling "Trig Out" Port
OUTP:EXT 1
Enable "Trig Out" port to output pulse
after each scanned channel is closed.
OUTPut[:EXTernal][:STATe]?
OUTPut[:EXTernal][:STATe]?
Queries the present state of the "Trig Out" port. The
command returns "1" if the port is enabled or "0" if disabled.
Example
Query "Trig Out" Port Enable State
This example enables the "Trig Out" port and queries the enable state. The
OUTPut?
command returns "1" since the port is enabled.
OUTP:EXT ON
Enable "Trig Out" port.
OUTP:EXT?
Query port enable state.
OUTPut:TTLTrgn[:STATe]
OUTPut:TTLTrg
n
[:STATe] <
mode
>
Selects and enables which TTL Trigger bus
line (0 to 7) will output a trigger when a channel is closed during a scan. This is also
used to disable a selected TTL Trigger bus line. "
n
" specifies the TTL Trigger bus
line (0 to 7) and "
mode
" enables
(
ON
or
1
) or disables (
OFF
or
0
) the specified TTL Trigger bus line.
Parameters
Comments
Enabling TTL Trigger Bus:
When enabled, a pulse is output from the selected
TTL Trigger bus line (0 to 7) after each channel in the switchbox is closed during a
scan. If disabled, a pulse is not output. The output is a negative-going pulse.
TTL Trigger Bus Line Shared by Switchboxes:
Only one switchbox
configuration can use the selected TTL Trigger at a time. When enabled, the selected
TTL Trigger bus line (0 to 7) is pulsed by the switchbox each time a scanned channel
is closed. To disable the output for a specific switchbox, send the
OUTPut:TTLTrg
n
OFF
or
0
command for that switchbox.
One Output Selected at a Time:
Only one output (
ECLTrg0
or
1
;
TTLTrg0
,
1
,
2
,
3
,
4
,
5
,
6
, or
7
; or
EXTernal
) can be enabled at one time. Enabling a different output
source will automatically disable the active output. For example, if
TTLTrg1
is the
active output and
TTLTrg4
is enabled,
TTLTrg1
will become disabled and
TTLTrg4
will become the active output.
Parameter
Name
Parameter
Type
Range of Values
Default
Value
n
numeric
0 to 7
N/A
mode
boolean
0 | 1 | OFF | ON
OFF|0
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...