86
Agilent E8462A Relay Multiplexer Command Reference
Related Commands:
[ROUTe:]SCAN
,
TRIGger:SOURce
,
OUTPut:TTLTrg
n
[:STATe]?
*RST Condition:
OUTPut:TTLTrg
n
[:STATe] OFF
(disabled)
Example
Enabling TTL Trigger Bus Line 7
OUTP:TTLT7:STAT 1
Enable TTL Trigger bus line 7 to output
pulse after each scanned channel is
closed.
OUTPut:TTLTrgn[:STATe]?
OUTPut:TTLTrg
n
[:STATe]?
Queries the present state of the specified TTL Trigger
bus line. The command returns "1" if the specified TTLTrg bus line is enabled or
"0" if disabled.
Example
Query TTL Trigger Bus Enable State
This example enables TTL Trigger bus line 7 and queries the enable state. The
OUTPut:TTLTrg
n
?
command returns "1" since the port is enabled.
OUTP:TTLT7:STAT 1
Enable TTL Trigger bus line 7.
OUTP:TTLT7?
Query bus enable state.