146
Error Messages
Table C-2. Multiplexer Error Messages
Code
Error Message
Potential Cause(s)
-211
Trigger ignored
Trigger received when scan not enabled. Trigger received after scan complete. Trigger
too fast.
-213
Init Ignored
Attempting to execute an INIT command when a scan is already in progress.
-222
Data out of range
Parameter value is outside valid range.
-224
Illegal parameter value
Attempting to execute a command with a parameter not applicable to the command.
-240
Hardware error
Command failed due to a hardware problem.
-310
System error
Internal driver error. This error can result if an excessively long parameter list is entered.
1500
External trigger source already
allocated
Assigning an external trigger source to a switchbox when the trigger source has already
been assigned to another switchbox.
1510
Trigger source
non-existent
Selected trigger source is not available on this platform (e.g. some triggers are not
available on the E1300/E1301 VXI B-size mainframes).
2000
Invalid card number
Addressing a module (card) in a switchbox that is not part of the switchbox.
2001
Invalid channel number
Attempting to address a channel of a module in a switchbox that is not supported by the
module (e.g., channel 99 of a multiplexer module).
2006
Command not supported on this
card
Sending a command to a module (card) in a switchbox that is unsupported by the
module.
2008
Scan list not initialized
Executing a scan without the INIT command.
2009
Too many channels in channel
list
Attempting to address more channels than available in the switchbox.
2010
Scan mode not allowed on this
card
The selected scanning mode is not allowed with this module or you have misspelled the
mode parameter (see SCAN:MODE command).
2011
Empty channel list
No valid channels are specified in the channel_list.
2012
Invalid Channel Range
Invalid channel(s) specified in SCAN <
channel_list
> command. Attempting to begin
scanning when no valid channel list is defined.
2600
Function not supported on this
card
Sending a command to a module (card) in a switchbox that is not supported by the
module or switchbox.
2601
Channel list required
Sending a command requiring a
channel_list
without the
channel_list
.
Содержание E8462A
Страница 2: ...x...
Страница 10: ...8 Notes...
Страница 11: ...9 Notes...
Страница 12: ...10 Notes...
Страница 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Страница 68: ...66 Using the Multiplexer...
Страница 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Страница 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Страница 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Страница 146: ...144 Register Based Programming...
Страница 156: ...154 Index...