Error Messages
4
User’s Guide
153
–330
Self-test Failed;RAM Battery Fault
Refer to
on page 98 if you require a description of the battery test.
–330
Self-test Failed;Calibrator Fault
Refer to
on page 99 if you require a description of the calibrator test.
–330
Self-test Failed;ROM Check Failed
Refer to
on page 98 if you require a description of the ROM Checksum test.
–330
Self-test Failed;RAM Check Failed
Refer to
on page 98 if you require a description of the RAM test.
–330
Self-test Failed;Display Assy. Fault
Refer to
on page 99 if you require a description of the Display test.
–330
Self-test Failed;Confidence Check Fault
Refer to
on page 96 if you require a description of this test.
–330
Self-test Failed;Serial Interface Fault
Refer to
on page 99 if you require a description of this test.
–350
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
–361
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be
guaranteed.
–362
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot
be guaranteed.
–363
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
–410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The output
buffer is cleared when power has been off or after
*RST
(reset) command has been
executed.
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...