134
E4418B Power Meter User’s Guide
•
Press this softkey to set the duty cycle for the pulse power measurement
feature of the power meter. A value in the range of 0.001% to 100% can be
entered. The default value is 1.000%. Use the
,
,
, and
keys to change the value. To confirm your choice press
.
is automatically set to “On” when a value is entered
using
on page 68 for
further information.
•
Press this softkey to access a menu which allows you to switch the filter
off and on, set the filter length, select automatic or manual mode and
switch step detection off and on.
■
Press this softkey to toggle the filter value on or off. The default is
“On”. The filter value is set using
■
Press this softkey to toggle between the automatic (AUTO) and
manual (MAN) filter modes. In manual mode, you enter the number of
measurements to be averaged. In automatic mode, the number of
measurements averaged is based on the measured power and uses a
pre- defined look- up table. Refer to
for further information.
■
Press this softkey to enter the filter length. The filter is used to reduce
noise, obtain the desired resolution and to reduce jitter in the
measurement
results.
Use
the , , ,
and
keys
to
change the value.
■
Press this softkey to switch step detection on or off. Step detection
reduces the settling time of the filter by re- initialising the filter after a
significant step (increase or decrease) has been detected in the
measured power. The default is “On”.
Duty Cycle
%
Duty Cycle Off On
Duty Cycle
Filter
Filter Off On
Length
Mode AUTO MAN
Length
Step Det Off On
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...