Error Messages
4
User’s Guide
147
Error queue messages have the following format:
For example,
–330, “Self-test Failed;Battery Fault”
.
Errors are retrieved in a first in first out (FIFO) order. If more than 30 errors
occur, the error queue overflows and the last error in the queue is replaced
with error
-350, “Queue Overflow”
. Any time the queue overflows, the
most recent error is discarded.
Once the errors are read, they are removed from the error queue. This opens a
position at the end of the queue for a new error message, if one is
subsequently detected. When all errors have been read from the queue, further
error queries return
+0, “No errors”
.
To delete all the errors in the queue from the front panel press:
•
,
,
then
.
To delete all the errors in the queue, remotely use:
•
the
*CLS
(clear status) command.
The error queue is also cleared when the instrument power has been switched
off.
Error
,
‚
‚
;
Device
Error
Description
Dependent Info
Number
Error List
Clear Errors
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...