4
Error Messages
152
E4418B Power Meter User’s Guide
–310
System error;Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected is for use with CW signals only.
–310
System error;Sensor EEPROM Read Failed - critical data not found or unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Sensor EEPROM Read Completed OK but optional data block(s) not found or
unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Sensor EEPROM Read Failed - unknown EEPROM table format
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp
(temperature compensation).
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Option 001 Battery charger fault
The power meter is connected to an AC power source, the battery is not fully charged and it is
not charging.
–310
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance, the power
meter detects power sensors connected to both its front and rear channel inputs.
–320
Out of memory
The power meter required more memory than was available to run an internal operation.
–330
Self-test Failed;
The –330, “Self-test Failed” errors indicate that you have a problem with your power meter.
Refer to
“Contacting Agilent Technologies”
on page 103 for details of what to do with your
faulty power meter.
–330
Self-test Failed;Measurement Channel Fault
Refer to
on page 98 if you require a description of the Measurement
Assembly test.
–330
Self-test Failed;Option 001 Battery requires replacement
The Option 001 battery is not charging to a satisfactory level and should be replaced.
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...